Features


Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top 

 

 Download Medical PIU Technical Brochure    medgroup5

 

 

Plitron Medical Power Isolation Units (PIUs) combine highly efficient toroidal isolation transformers, series-mode surge suppression, and certification to international safety standards IEC/CSA/UL 60601-1, to provide the ultimate in power line protection and low-leakage current. Plitron Medical PIUs use patented series-mode surge removal filters to absorb dangerous voltage surges up to 6000 volts and safely dissipate them without using failure-prone MOVs (metal oxide varistors). Surges are fed onto the line, not shunted to ground, ensuring the line remains clean of contamination. Plitron Medical PIUs are available in power ranges of 300, 600, 900, 1200, 1800 and 2400VA in both North American (120V) or Universal (100-120-230/240V) operations. Attractive and rugged metal cases house the units.  

 

 

Specifications


Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top

 

 

 

Universal Units

North American Units

Input  VAC +/- 10%

100-120-230/240VAC selectable, 50/60Hz.

120VAC, 60 Hz

Inputs

IEC Inlet

Plug and line cord: Hubbell hospital-grade, grounded, 3-wire, 6’ long

Isolation

Primary to Secondary4,000 VAC, primary to secondary. 100% hi-pot tested for 1 minute

Input Protection

Dual fuses, panel-mount

Circuit breaker, resettable, panel-mount

Outputs

IEC outlets

Receptacles: Duplex, Hubbell medical-grade.

Surge Suppression

(Up to 240Vrms): Surges of up to 6000volts (1.2 X 50 USec industry standard rating), with the clamping voltage onset of 344volts peak

(120Vrms): Surges of up to 6000volts (1.2 X 50μSec industry standard rating) with the clamping voltage onset of 172volts peak.

 

Universal Models


MODEL

Input Rating

Current A / Input V

Output Rating

Load
Reg.

Leakage Current

Size (cm)

Weight

100

120

230

240

100

120

230

240

%

(Note 1)

W

H

L

kg

PM-09236-00

3.0

2.5

1.3

1.25

3.0

2.5

1.3

1.25

4.2

30

14.74

11.6

24.7

10.22

PM-09237-00

6.0

5.0

2.6

2.5

6.0

5.0

2.6

2.5

4.0

50

19.8

11.9

27.0

14.0

PM-09238-00

9.0

7.5

3.9

3.75A

9.0

7.5

3.9

3.75

3.5

70

24.1

15.0

31.8

21.3

PM-09239-00

10.0

10.0

5.2

5.0

9.0

9.0

5.2

5.0

3.0

70

24.1

15.0

31.8

24.5

PM-09240-00

15.0

15.0

7.8

7.5

15.0

15.0

7.8

7.5

2.0

80

31.0

16.8

38.8

39.0

PM-09241-00

16.0

16.0

10.4

10.0

16.0

16.0

10.4

10.0

2.0

90

31.0

16.8

38.8

45.9

Note 1 – Typical, Measured Ua at 110% Rated Voltage

 

North American Models


MODEL

Input  IAC

Output Rating

Load
Regulation

Leakage Current

Size (cm)

Weight

@ 120V

120V

%

(Note 1)

W

H

L

kg

PM-08905-00

2.5

2.5

4.50

30

14.74

11.6

24.7

9.3

PM-08906-00

5.0

5.0

5.00

40

19.8

11.9

27.0

12.5

PM-08983-00

7.5

7.5

5.00

40

24.1

15.0

31.8

17.5

PM-08907-00

10.0

10.0

5.00

40

24.1

15.0

31.8

20.5

PM-08908-00

15.0

15.0

5.00

40

31.0

16.8

38.8

30.7

PM-08909-00

16.0

16.0

5.00

50

31.0

16.8

38.8

36.7

Note 1 – Typical, Measured UA at 110% Rated Voltage

 

Rear Panel Layouts


Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top

 

Click On Image To Open in New Window ' + ' to Enlarge

 

 

Regulatory Approvals


Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top

 

  • CAN/CSA C22.2 No 0-M91 – General Requirement Canadian Electrical Code, Part II

  • CAN/CSA C22.2 No 601.1 – M90 – Medical Electrical Equipment Part I : General

  • CAN/CSA C22.2 No 601.1S1 – 94 – Supplement No. 1-94 to 601.1 – M90

  • CSA 601.1 Amendment 2:1998

  • UL 60601-1 – :1998 –, including Amendment 1 (1991) and Amendment 2 (1995)

  • Medical Electrical Equipment Part I : General Requirement for Safety

 

Schematics


Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top

 

  • North American 120 VAC Input, 60 Hz

  • Universal Version, 100, 120, 230, 240 VAC Input, 0-60 Hz

Features | Specs | Mechanical | Regulatory Approvals | Schematics | Top