POWER ISOLATION FINISHED UNITS

Features

  • International approvals, IEC/CSA/UL 60601-1
  • Low leakage current (typically <30uA)
  • Surge suppression (handles surges up to 6000 volts
  • All units feature Plitron toroidal isolation transformers
  • 12 models in power ranges from 300 to 2400VA in North American and Universal voltages

Approvals

  • CAN/CSA C22.2 No 0-M91 – General Requirement Canadian Electrical Code, Part II
  • CAN/CSA C22.2 No 601.1 – M90 – Medical Electrical Equipment Part I : General
  • CAN/CSA C22.2 No 601.1S1 – 94 – Supplement No. 1-94 to 601.1 – M90
  • CSA 601.1 Amendment 2:1998
  • UL 60601-1 – :1998 –, including Amendment 1 (1991) and Amendment 2 (1995)
  • Medical Electrical Equipment Part I : General Requirement for Safety

Plitron Medical Power Isolation Units (PIUs) combine highly efficient toroidal isolation transformers, series-mode surge suppression, and certification to international safety standards IEC/CSA/UL 60601-1, to provide the ultimate in power line protection and low-leakage current.

Plitron Medical PIUs use patented series-mode surge removal filters to absorb dangerous voltage surges up to 6000 volts and safely dissipate them without using failure-prone MOVs (metal oxide varistors).

Surges are fed onto the line, not shunted to ground, ensuring the line remains clean of contamination. Plitron Medical PIUs are available in power ranges of 300, 600, 900, 1200, 1800 and 2400VA in both North American (120V) or Universal (100-120-230/240V) operations. Attractive and rugged metal cases house the units.

Schematic Diagram North American Models

Schematic Diagram Global Models

Universal Units North American Units
Input  VAC +/- 10% 100-120-230/240VAC selectable, 50/60Hz. 120VAC, 60 Hz
Inputs IEC Inlet Plug and line cord: Hubbell hospital-grade, grounded, 3-wire, 6’ long
Isolation Primary to Secondary4,000 VAC, primary to secondary. 100% hi-pot tested for 1 minute
Input Protection Dual fuses, panel-mount Circuit breaker, resettable, panel-mount
Outputs IEC outlets Receptacles: Duplex, Hubbell medical-grade.
Surge Suppression (Up to 240Vrms): Surges of up to 6000volts (1.2 X 50 USec industry standard rating), with the clamping voltage onset of 344volts peak (120Vrms): Surges of up to 6000volts (1.2 X 50μSec industry standard rating) with the clamping voltage onset of 172volts peak.
MODEL Input  IAC Output Rating Load
Regulation
Leakage Current Size (cm) Weight
@ 120V 120V % (Note 1) W H L kg
PM-08905-00 2.5 2.5 4.50 30 14.74 11.6 24.7 9.3
PM-08906-00 5.0 5.0 5.00 40 19.8 11.9 27.0 12.5
PM-08983-00 7.5 7.5 5.00 40 24.1 15.0 31.8 17.5
PM-08907-00 10.0 10.0 5.00 40 24.1 15.0 31.8 20.5
PM-08908-00 15.0 15.0 5.00 40 31.0 16.8 38.8 30.7
PM-08909-00 16.0 16.0 5.00 50 31.0 16.8 38.8 36.7
Note 1 – Typical, Measured UA at 110% Rated Voltage
Input Rating
Current A / Input V
Output Rating Load
Reg.
Leakage Current Size (cm) Weight
100 120 230 240 100 120 230 240 % (Note 1) W H L kg
PM-09236-00 3.0 2.5 1.3 1.25 3.0 2.5 1.3 1.25 4.2 30 14.74 11.6 24.7 10.22
PM-09237-00 6.0 5.0 2.6 2.5 6.0 5.0 2.6 2.5 4.0 50 19.8 11.9 27.0 14.0
PM-09238-00 9.0 7.5 3.9 3.75A 9.0 7.5 3.9 3.75 3.5 70 24.1 15.0 31.8 21.3
PM-09239-00 10.0 10.0 5.2 5.0 9.0 9.0 5.2 5.0 3.0 70 24.1 15.0 31.8 24.5
PM-09240-00 15.0 15.0 7.8 7.5 15.0 15.0 7.8 7.5 2.0 80 31.0 16.8 38.8 39.0
PM-09241-00 16.0 16.0 10.4 10.0 16.0 16.0 10.4 10.0 2.0 90 31.0 16.8 38.8 45.9
Note 1 – Typical, Measured Ua at 110% Rated Voltage